Maintenance points of several types of incubators

Digital display constant temperature shaking incubator maintenance:
1. Accurate use and attention to the maintenance of the instrument, so that it is in good working condition, can extend the life of the instrument.
2. During continuous operation, the instrument should perform a periodic inspection every three months; check the fuses, control components and fastening screws, and whether there are water droplets, dirt, etc. falling into the motor and control components.
3. The bearing of the transmission part has been filled with the appropriate amount of grease (No. 1 calcium-nano) before leaving the factory. During the continuous operation, the instrument should be filled with grease once every six months. The filling amount is about 1/3 of the bearing space. .
4. The instrument is often used, and natural wear is normal. After one year of use, if the motor or compressor is found to have abnormal noise; the transmission part bearing is worn, the belt is loose or cracked, and the electronic control component fails, the company will inherit and provide quality service to assist in handling.
5. The compressor should avoid continuous starting.

Digital biochemical incubator maintenance and maintenance:
1 Observe the temperature of the incubator daily to meet the requirements. If you find an abnormal situation, please ask a professional to repair it.
2 When the lighting in the box is abnormal, it is necessary to replace the lighting fixtures in time.
3 Check if the heating wire (compressor) is damaged when the heating (cooling) is abnormal, and repair it in time after damage.
4 Clean the outside of the incubator in time to keep it clean and dust-free. Thoroughly clean the incubator liner once a week with sterile water. Replace different types of disinfectant water every quarter to avoid the emergence of resistant strains.
5 A performance verification of this equipment should be performed annually.
6 Before using the device, carefully check that the power supply voltage matches the instrument requirements.
7 When the device is not in use, set each control switch to a non-operational state and turn off the power.

Maintenance and maintenance of the digital light incubator:
1. The incubator housing should be reliably grounded.
2. The incubator should be placed in a cool, dry, well ventilated place away from heat and sunlight. Place it smoothly to prevent noise from occurring.
3. In order to ensure the effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm, the side of the box should have a gap of 50mm, and the top of the box should have at least 300mm space.
4. When the incubator is transported, repaired and maintained, it should avoid collision, shaking and shaking, and the maximum inclination is less than 45 degrees.
5. The instrument suddenly does not work. Please check if the fuse tube (after the box) is burnt out and check the power supply.
6. When the incubator is working in refrigeration, it is not advisable to make the difference between the temperature inside the box and the ambient temperature greater than 25 degrees.
7. The illumination time can be programmed and controlled by the “programmable time controller”. Please refer to the "Programmable Time Controller" instructions for programming methods.

Daily maintenance of water jacketed carbon dioxide incubator:
Since such instruments are often in contact with water and gas, impurities in the gas may enter the gas pipeline, causing the machine to malfunction. In daily work, it is necessary to carry out maintenance from the following aspects: always check the water level of the water jacket, if the water level is low, it is necessary to add water; check the carbon dioxide cylinder regularly, do not let the carbon dioxide cylinder empty; check the carbon dioxide gas supply pipeline and interface for leaks Gas phenomenon; regularly dust the machine to prevent dust from blocking the air passage and solenoid valve.
Note:
1. For initial use, be sure to add sufficient deionized or distilled water and cover with a sealing cap to reduce the evaporation of water in the water jacket.
2. In units with unstable power supply, it is best to equip them with high-performance voltage regulators to reduce faults caused by voltage instability.
3. Install a carbon dioxide pressure reducing valve with a low pressure of 0~0.25Mpa on the carbon dioxide cylinder to make the output pressure stable, so that the pressure of carbon dioxide in the input box can be avoided. The CO2 control valve is damaged during long-term working conditions.
5. When the power is not used for cultivation, the temperature should be set to 00C, and the carbon dioxide concentration should be set to “0”, so that the control valves are not in operation, thus avoiding the long-term control valve. Damaged by working condition.
7. The temperature of the door should be set higher than the temperature inside the box by about 10C to reduce the fluctuation of the temperature inside the box when the item is accessed.

Maintenance of the shaker incubator:
1. During the continuous operation, the instrument should do a regular inspection every three months; check the fuses, control components and fastening screws, and whether there are water droplets, dirt and other falling into the motor and control components.
2. The bearing of the transmission part has been filled with the proper amount of grease before leaving the factory. During the continuous operation, the instrument should be filled with grease once every six months, and the filling amount accounts for 1/3 of the bearing space.
3. The compressor should avoid continuous starting
4, the instrument is used frequently, natural wear is a normal phenomenon. After one year of use, if the motor or compressor is found to have abnormal noise; the transmission part bearing is worn, the belt is loose or cracked, and the electronic control component fails, the company will continue to provide quality service and assist in handling it.

Bulk of varicella vaccine,semi-Finished Products.It has good safety with low ratio in adverse reaction. And it has been exported to Bangladesh. 

In 2008, BCHT successfully launched its product Varicella Vaccine, Live in China. In the following years, BCHT continuously committed to improving the product and led in 2010 removal of gelatin from adjuvant and extended the vaccine shelf-life up to 36 months which is the longest one in the world in 2011. And it has the following qualities.


Varicella Vaccine

Final Bulk

Rabies Vaccine,Freeze-Dried Rabies Vaccine,Mild Rabies Vaccine,Stable Rabies Vaccine For Human Use

Changchun BCHT Biotechnology Co.,Ltd , https://www.ccbcht.net